-de-la-methode-operatoire-a-laudit">Retour au Guide Complet ISO 17025The calibration of a thermometer covering the range -196°C to +200°C involves two distinct physical domains: the cryogenic domain and the high temperature domain, in accordance with Guide FD X 07-011.
1. Reference Equipment by Range
| Range | Reference Equipment | Calibration Medium |
|---|---|---|
| -196°C to -100°C | SPRT Thermometer | Controlled liquid nitrogen bath |
| -100°C to 0°C | Pt100 Thermometer | Alcohol cryostatic bath |
| 0°C to +100°C | Pt100 Thermometer | Water-controlled bath |
| +100°C to +200°C | Pt100 Thermometer | Silicone oil-controlled bath |
2. Control Points | Variable MPE
| Nominal Point | Average | Deviation | MPE | Status |
|---|---|---|---|---|
| -196.0°C | -195.67 | +0.33 | ±0.5°C | ✅ CONFORMING |
| -100.0°C | -99.73 | +0.27 | ±0.5°C | ✅ CONFORMING |
| -10.0°C | -9.77 | +0.23 | ±0.5°C | ✅ CONFORMING |
| 0.0°C | +0.23 | +0.23 | ±0.5°C | ✅ CONFORMING |
| +50.0°C | +50.37 | +0.37 | ±0.5°C | ✅ CONFORMING |
| +100.0°C | +100.57 | +0.57 | ±1.0°C | ✅ CONFORMING |
| +200.0°C | +200.77 | +0.77 | ±1.0°C | ✅ CONFORMING |
3. Thermal Variation Correction
A. Spatial inhomogeneity of the bath (u_homo)
u_homo = ΔT_spatial / (2√3) = 0.1 / (2√3) = 0.029°C
B. Temporal instability of the bath (u_stab)
u_stab = ΔT_temporal / (2√3) = 0.05 / (2√3) = 0.014°C
4. Uncertainty Budget
| Source | Value (k=1) |
|---|---|
| u_homo | 0.029°C |
| u_stab | 0.014°C |
| u_ref | 0.050°C |
| u_res | 0.029°C |
| u_drift | 0.020°C |
| Expanded U (k=2) | 0.14°C |
ILAC-G8 at +200°C point: |0.77| + 0.14 = 0.91°C < MPE 1.0°C → ✅ CONFORMING
5. ISO 17025 Compliance
- Traceability: Traceability chain to LNE
- §7.6: Expanded uncertainty U declared on the PV for each range
- §7.8.6: ILAC-G8 rule applied point by point
- Periodicity: 12 months — extendable to 24 months via OIML D10